Case study · Analytical and Bioanalytical Support

Development and Evaluation (Quantitation) of the Anti-Drug Antibodies (ADA) Assay in the Dog Plasma

Methods

The purpose of this study was to develop and qualify a method for detection of anti-drug antibodies (ADA) against therapeutic peptide with nanocarrier in dog plasma using a Meso Scale Diagnostics (MSD) Electrochemiluminescence Immunoassay. The qualification plan was approved by the Sponsor. MSD utilizes electrochemiluminescence detection to detect binding events on patterned arrays by converting electric energy into light energy. Electrochemiluminescence detection offers a unique combination of sensitivity, dynamic range and convenience.

The MSD assay measures ADA against the therapeutic nanomaterial in dog plasma in an MSD 96-well-standard plate. In brief, the assay employs a sandwich immunoassay format where wells were coated with the drug on the standard MSD plate containing high-binding electrodes. Samples in serial dilutions and a solution containing the labeled secondary detection antibody are added to the plate. The ADA in the sample binds to the drugs immobilized on the working electrode surface; recruitment of the labeled detection antibody to the ADA completes the sandwich. Then an MSD read buffer is added to the plate that provides the appropriate chemical environment for electrochemiluminescence and the plate is loaded into an MSD SECTOR instrument for the analysis, see Figure 1.

Electrochemiluminescence sandwich immunoassay, antibody capture Diagram of one well. Capture antibodies are coated on the working electrode. One binds the analyte, and a detection antibody carrying a SULFO-TAG label binds the analyte from above. With read buffer in the well and voltage applied to the electrode, the label emits light. Read buffer Voltage applied Light SULFO-TAG label Detection antibody Analyte Capture antibody Working electrode
Electrochemiluminescence sandwich immunoassay, antibody capture Diagram of one well. Capture antibodies are coated on the working electrode. One binds the analyte, and a detection antibody carrying a SULFO-TAG label binds the analyte from above. With read buffer in the well and voltage applied to the electrode, the label emits light. Read buffer Voltage applied Light SULFO-TAG label Detection antibody Analyte Capture antibody Working electrode
Figure 1. MSD detection using electrochemiluminescence (antibody capture is shown, in this assay drug capture was used).

The ADA from dog plasma will be detected using the immobilized Test Article (drug) for capture with anti-dog secondary antibody linked with sulfo-TAG streptavidin for development of the assay. The system must be qualified for different species (i.e. rat, dog, human). In order to qualify the ADA assay, we designed an assay utilizing ADA raised in rabbits immunized with the drug (the only available specific antibody to the drug), see Figure 2.

Qualification format: coated drug, rabbit anti-drug antibody, anti-rabbit secondary antibody Diagram of one well. The drug (test article) is coated on the working electrode. A rabbit polyclonal anti-drug antibody binds the coated drug, and an anti-rabbit secondary antibody carrying a SULFO-TAG label binds the rabbit antibody. With read buffer in the well and voltage applied to the electrode, the label emits light. Read buffer Voltage applied Light SULFO-TAG label Anti-rabbit secondaryantibody Rabbit polyclonalanti-drug antibody Coated drug (test article) Working electrode
Qualification format: coated drug, rabbit anti-drug antibody, anti-rabbit secondary antibody Diagram of one well. The drug (test article) is coated on the working electrode. A rabbit polyclonal anti-drug antibody binds the coated drug, and an anti-rabbit secondary antibody carrying a SULFO-TAG label binds the rabbit antibody. With read buffer in the well and voltage applied to the electrode, the label emits light. Read buffer Voltage applied Light SULFO-TAG label Anti-rabbitsecondary antibody Rabbit polyclonalanti-drug antibody Coated drug(test article) Working electrode
Figure 2. MSD detection using coated drug for capture, shown with rabbit polyclonal anti-drug Antibody and anti-rabbit secondary antibody, used for development and qualification of the assay.

Results

The method was qualified for accurately measuring induction of Ab to the test article in dog plasma. The screen cut-off point was determined for each dilution of plasma. The data showed good precision (%CV) and good accuracy (%RE) for both intra- and inter-assay calculations confirming acceptable selectivity of the assay.

The range of quantitation during sample analysis was 0.49 ng/L – 500 ng/mL. The limit of detection was 0.456 ng/mL. The lowest limit of qualification (LLOQ) was 0.98 ng/mL. The minimum required dilution (MRD) of samples was 1:50 (2% plasma) for this assay. This allow to use as low as 2.5 uL of plasma for ADA analysis. Using of this method confirmed that there were no ADA induced in dogs due to the drug administration.